ISO-PUCK

Overview

Electrical Safety Standards to prevent Electric Shock

Over the past several years we have seen an increased national and international awareness of the potential hazard to humans of electrical shock from patient care equipment. This concern produced the third edition of the safety standard IEC60601-1:2005. In the US the equivalent safety standard is ANSI/AAMI ES60601-1:2005(R) 2012

ISO-PUCK medical grade isolation transformers prevent hazardous and possibly life threatening levels of leakage ( touch) current in medical devices and commercial grade electronics including computers, monitors, printers and other non-battery operated electrical equipment used in OR and patient care areas.

ISO-PUCK isolation transformers do not feature output Neutral to Ground Bond. They provide a true floating system (IT Earthing System) for maximum protection against excessive ground leakage current.

In the United States, the 2012 edition of NFPA 99 went through a major overhaul as it was changed from a standard to a code to reflect how the document is used and adopted. It is now a risk based document.  Maximum allowed level of leakage (touch) current is 100 micro amps in fixed equipment and 500 micro amps in portable equipment.

 

Shop Now

Over the past several years we have seen an increased national and international awareness of the potential hazard to humans of electrical shock from patient care equipment. This concern produced the third edition of the safety standard IEC60601-1:2005. In the US the equivalent safety standard is ANSI/AAMI ES60601-1:2005(R) 2012

ISO-PUCK medical grade isolation transformers prevent hazardous and possibly life threatening levels of leakage ( touch) current in medical devices and commercial grade electronics including computers, monitors, printers and other non-battery operated electrical equipment used in OR and patient care areas.

ISO-PUCK isolation transformers do not feature output Neutral to Ground Bond. They provide a true floating system (IT Earthing System) for maximum protection against excessive ground leakage current.

In the United States, the 2012 edition of NFPA 99 went through a major overhaul as it was changed from a standard to a code to reflect how the document is used and adopted. It is now a risk based document.  Maximum allowed level of leakage (touch) current is 100 micro amps in fixed equipment and 500 micro amps in portable equipment.

 

 

Power Strips (RPT’s Relocatable Power Taps)

NFPA 99 does not specify the safety standard to which a power strip (RPT) must be listed. There are two standards applicable to power strips in healthcare facilities. UL1363A for use in patient care areas and UL1363 for non-patient care areas.

Unless otherwise instructed by our customer we will always provide our ISO-PUCK with a power strip listed to UL1363A.

Wet Locations

Whenever an operating room is designated as a wet location, installing an isolation transformer like the ISO-PUCK rather than a ground fault circuit interrupter (GFCI) avoids interruption of power during surgery.

 

ISO-PUCK Models for 120V input and 120V output. Voltage at 60 Hz Frequency

Power RatingModelHousing Material* (Finish)Mounting                    SEE DIM ADiameter – SEE DIM CDepth SEE DIM BApprox. Weight
In.mmIn.mmLbsKg
200 VAIP-200/60Polycarbonate (White)5.5in (140mm)4 7/81252 1/8545.12.3
300 VAIP-300/60Polycarbonate (White)6 5/8in (169mm)5 3/41452 3/4707.94.0
400 VAIP-400/60Polycarbonate (White)6 5/8in (169mm)5 3/41452 3/4708.03.63
600 VAIP-600/60Polycarbonate (White)8in (203.5mm)6 5/81682 3/47012.05.45
800 VAIP-800/60Polycarbonate (White)8in (203.5mm)6 5/81682 3/47014.06.4
1000 VAIP-1000/60Polycarbonate (White)9 3/8in (238mm)7 7/82003 1/88019.78.9
1200 VAIP-1200/60Polycarbonate (White)9 3/8in (238mm)7 7/82003 1/880209.1
1500 VAIP-1500/60Polycarbonate (White)9 7/8in (249mm)8 1/22183 3/41002812.7
1800 VAIP-1800/60Polycarbonate (White)9 7/8in (249mm)8 1/22183 3/41003013.6
2200 VAIP-2200/60Polycarbonate (White)9 7/8in (249mm)8 1/22183 3/41003315.0
2400 VAIP-2400/60Enamel Coated Steel (Off-White)Single Screw Mounting9   2304 1/41083515.9
3000 VAIP-3000/60Enamel Coated Steel (Off-White)Single Screw Mounting9   2304 1/41083817.2
*Certain models with Polycarbonate housings  are available in Enamel Coated Steel, Please call factory for details

 

ISO-PUCK Models for 100V or 120V or 230/240V. Input or Output Voltages at 50-60 Hz Frequency

Power RatingModelHousing Material* (Finish)Mounting                    SEE DIM ADiameter – SEE DIM CDepth – SEE DIM BApprox. Weight
In.mmIn.mmLbsKg
200 VAIP-200/50Polycarbonate (White)5.5in (140mm)4 7/81252 1/8546.22.8
350 VAIP-350/50Polycarbonate (White)6 5/8in (169mm)5 3/41452 3/47013.36.0
650 VAIP-650/50Polycarbonate (White)8in (203.5mm)6 5/81682 3/47017.47.9
1000 VAIP-1000/50Polycarbonate (White)9 3/8in (238mm)7 7/82003 1/880209.1
1200 VAIP-1200/50Polycarbonate (White)9 3/8in (238mm)7 7/82003 1/8802611.8
1800 VAIP-1800/50Polycarbonate (White)9 7/8in (249mm)8 1/22183 3/41003315.0
2500 VAIP-2500/50Enamel Coated Steel (Off-White)Single Screw Mounting9   2304 1/41083817.2
*Certain models with Polycarbonate housings  are available in Enamel Coated Steel, Please call factory for details

 

 Plugs and Receptacles

USA/Canada:
For 120V/15A line voltage NEMA 5-15 Hospital Grade
For 120V/20A line voltage NEMA 5-20 Hospital Grade
For higher amperage please consult factory. In addition to NEMA 5 receptacle,
a triple outlet and a four or six outlet hospital grade power strip permanently attached
are available.

Europe: Schuko, UK or as specified.
Rest of the world: As specified.

IEC60320 connectors are available for all models.

Cord Lenghts and Color:
Standard length is 2 feet (60 cm). Custom lengths available at nominal cost.
White color is standard.

 

ISO-Tower – New Product

  •  PORTABLE MULTI-TAP ISOLATED POWER SOLUTION
  •  CENTRALLY LOCATED TO MINIMIZE MULTIPLE CABLING ISSUES
  •  1200VA ISOLATION TRANSFORMER WITH 6 GREEN-DOT OUTLETS AND A 10FT POWER CORD
  •  CERTIFIED TO  UL IEC60601.1 THIRD EDITION

 

Medical Safety Standard Certifications

Our ISO-PUCK series of enclosed, corded hospital grade isolation transformers are certified to safety standards.
ANSI/AAMI (formerly UL) ES60601-1, CSA std c22.2, No.601.1, EN 61558-2, and IEC 60601-1 third edition.
Our certifications to these safety standards also include our TIS – series bare medical grade transformers.
As of June 1, 2013 the third edition is mandatory for all new medical devices to be used in the health care industry in North America, Europe and countries requiring compliance to IEC 60601.1.  
Our transformers carry  CE marking and comply with current RoHS requirements.

Intertek File: 3051384.

CB Test Report: Reference No. 100891440BOX-002 issued March 26, 2013 

BENEFITS OF USING CERTIFIED MEDICAL GRADE ISOLATION TRANSFORMERS
Adding a medical grade isolation transformer to your product avoids expensive and time consuming safety testing bringing your product to the market much sooner.

Download ISO-Puck Brochure

 

Customer List

Health Care Facilities
 Medical Equipment Manufacturers
Adventist Medical Center
Blackrock NeuroMed
Carolinas Healthcare System
Dartmouth-Hitchcock Medical Center
Feinstein Institute, Medical Neuroanatomy
Florida Hospital, Nicholson Center
GE Healthcare
Halifax Regional Medical Center
Iredell Memorial Hospital
Kaiser Permanente
Memorial Hospital, Gulfport
St. Luke’s University Health Network
Tampa General Hospital
University of Illinois
Cardiovascular Engineering, Inc.
Caliber Imaging and Diagnostics, Inc.
Carl Zeiss, Canada
3M Oral Care
ConMed New York
ConMed Australia
FotoFinder Systems, Inc
Image Diagnostics (Mass)
iVu Imaging Corporation
Kubtec, Inc.
Medtronics
Mevion Medical Systems
Olympus America, Inc.
OraMetrix, Inc. Texas
Ora-Metrix GmbH, Germany
Ortho Cindematics
Pentax Medical
Smart Tech Imaging

 

Reach Us

Contact Us for Information and Sales

6 Waterview Drive, Shelton, CT 06484

sales@bridgeportmagnetics.com

    Contact